Model Number FINETUNER ECHO [FINETUNER 2] |
Device Problems
No Device Output (1435); Output Problem (3005)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
|
Event Description
|
A finetuner echo was returned to service and repair with the complaint of a low battery life.
|
|
Event Description
|
A finetuner echo was returned to service and repair with the complaint of a low battery life.The device had been in use since december 2021.
|
|
Manufacturer Narrative
|
Conclusion: according the information received, a finetuner echo was sent to service _ repair due to a low battery life.During device investigation a defective capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.No injury was reported.This is a final report.
|
|