DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR AND POWER IV; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that unit has a bad power outlet module.It was running the old software.There were signs of melting on the module.The event timing was outside of surgery.There was no harm or delay reported.Upon investigation, the power inlet module was burned.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the power inlet module was burned.The power inlet module was replaced and resolved the reported issue.The root cause of the reported issue is attributed to the time the high flow valve being left open being too short during the pump start up sequence, a lack of clarity in defining required approvers for design related test protocols, and the servicing defect does not require trending.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|