• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR LIMITED NUCLEUS CI512 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR LIMITED NUCLEUS CI512 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI512
Device Problem No Device Output (1435)
Patient Problems Head Injury (1879); Failure of Implant (1924)
Event Date 11/25/2022
Event Type  malfunction  
Manufacturer Narrative
This report is submitted on december 16, 2022.
 
Event Description
Per the clinic, the patient experienced loss of connection to the internal device subsequent to sustaining a head trauma twice in august 2022 (specific dates not reported).Hardware exchange attempts were made; however, the issue could not be resolved.The implanted device remains.Explant and reimplantation is planned but has not taken place at the time of this report.
 
Manufacturer Narrative
Per the clinic, the device was explanted on (b)(6)2023 and was reimplanted with another cochlear device during the same surgery.This report is submitted on february 20, 2023.
 
Manufacturer Narrative
Device analysis indicated device failure.This report is submitted on march 14, 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI512 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
ken yian chow
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key15992909
MDR Text Key305584436
Report Number6000034-2022-03789
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502014564
UDI-Public(01)09321502014564(11)181213(17)201212
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/12/2020
Device Model NumberCI512
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/25/2022
Initial Date FDA Received12/15/2022
Supplement Dates Manufacturer Received01/30/2023
03/09/2023
Supplement Dates FDA Received02/20/2023
03/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age4 YR
Patient SexMale
-
-