MEDOS INTERNATIONAL SARL FMS TORNADO MICRO II NO-BUTTON SHAVER HANDPIECE; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283712
Device Problems Mechanical Problem (1384); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).Reporter is a j&j sales representative.Investigation summary: the complaint device was received at the r&d & manufacturing site and evaluated.During the analysis of the device, it was deemed that the motor presented a failure; however, as the motor cannot be disassembled to perform a repair, the devices must be discarded.The front seal of the motor is allowing saline to enter the motor during use.This causes the handpiece to stop functioning.This fault affects the reliability of the handpiece but does not adversely impact patient safety.This issue has been escalated to capa.A manufacturing record evaluation was performed for the finished device serial number (b)(4), and no non-conformance was identified.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
Shaver no patient harm or case delay.It was reported by the sales rep that during a shoulder repair surgery on (b)(6) 2022, it was observed that the fms tornado micro ii no-button shaver handpiece device made loud noise, would not oscillate and made the machine beep.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS TORNADO MICRO II NO-BUTTON SHAVER HANDPIECE
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15996098
MDR Text Key308236284
Report Number1221934-2022-04003
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705030736
UDI-Public10886705030736
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283712
Device Catalogue Number283712
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/23/2022
Initial Date FDA Received12/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2022
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-