• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2022
Event Type  malfunction  
Manufacturer Narrative
Patient sample: (b)(6) was requested for further investigation.This patient sample will be investigated for possible interfering factors.
 
Event Description
The initial reporter received questionable elecsys tsh ver.2 assay, elecsys ft4 iii assay, elecsys ft3 iii and elecsys ft3 iii ver.2 results for 2 patient samples tested on the customer's cobas e 801 module with serial number: (b)(4), the investigation site's cobas e 411 module with serial number: (b)(4) and the investigation site's cobas e 801 module with serial number: (b)(4).This medwatch is for the ft4 iii assay.Refer: to medwatch with patient identifier: (b)(6) for the ft3 iii assay; medwatch with patient identifier: (b)(6) for the ft3 iii v2 assay; medwatch with patient identifier: (b)(6) for the tsh v2 assay.The exact date of the event is unknown.The date the information was received was used as the date of the event.The initial results were not reported outside of the laboratory.The patient samples were sent to an investigation site that uses an e 411 and e 801.It was also sent to an outsourced laboratory that uses an abbott architect.The ft4 iii reagent lot number used at the customer site's e 801 was requested but not provided.The ft4 iii reagent lot number used at the investigation site's e 411 is 593219 with an expiration date of 31-dec-2022.The ft4 iii reagent lot number used at the investigation site's e 801 is 630888 with an expiration date of 31-may-2023.Refer to the attachment in the medwatch for the highlighted questionable results.
 
Manufacturer Narrative
The qc and calibration generated at the investigation site were within specifications.The cause of the event could not be determined as the patient samples were not available for further investigation.Product labeling states: "for diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings." medwatch field b6 - relevant test/laboratory data was updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15996810
MDR Text Key308250413
Report Number1823260-2022-04051
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976836190
Device Lot NumberASKU, 593219, 630888
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/21/2022
Initial Date FDA Received12/16/2022
Supplement Dates Manufacturer Received02/13/2023
Supplement Dates FDA Received03/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-