• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC PEG 24 PERCUTANEOUS ENDOSCOPIC GASTROSTOMY SET - PULL; KNT, TUBE, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC PEG 24 PERCUTANEOUS ENDOSCOPIC GASTROSTOMY SET - PULL; KNT, TUBE, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Catalog Number PEG-24-PULL-I-S
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abdominal Distention (2601); Increased Intra-Peritoneal Volume (IIPV) (4498)
Event Date 11/23/2022
Event Type  Injury  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The complications reported are known clinical complications.There is no evidence to suggest the report is due to a device failure.The ifu states the following potential complications: ¿potential complications associated with placement and use of a peg tube include, but are not limited to: bronchopulmonary aspiration and pneumonia, respiratory distress or airway obstruction, peritonitis or septic shock, colocutaneous, gastrocolocutaneous or small bowl fistula, gastric dilation, sigmoid intra-abdominal herniation and volvulus, persistent fistula following peg removal, esophageal injury, necrotizing fasciitis, candida cellulitis, improper placement or inability to place peg tube, tube dislodgement or migration, hemorrhage, and tumor metastasis.Additional complications include, but are not limited to: pneumoperitoneum, peristomal wound infection and purulent drainage, stomal leakage, bowel obstruction, gastroesophageal reflux (gerd), and blockage or deterioration of the peg tube." prior to distribution, all percutaneous endoscopic gastrostomy sets are subjected to a visual and functional inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During a gastronomy feeding tube placement, the physician used a cook peg 24 percutaneous endoscopic gastrostomy set - pull.It was reported [that seven days after tube placement] patient has fever, slight redness and swelling of the upper abdominal wall, [and] subcutaneous touch hairtwist sensation in the upper abdomen.Abdominal ct showed that the gastrostomy tube protruded subcutaneously and subcutaneous accumulation of fluid and gas in the upper abdominal wall.A section of the device did not remain inside the patient's body.The patient experienced fever, slight redness, and swelling of the upper abdominal wall and required emergency room visit due to this occurrence.Emergency treatment included subcutaneous exploration of the abdominal wall and debridement to place gastronomy tube.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEG 24 PERCUTANEOUS ENDOSCOPIC GASTROSTOMY SET - PULL
Type of Device
KNT, TUBE, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key15996963
MDR Text Key305623041
Report Number1037905-2022-00685
Device Sequence Number1
Product Code KNT
UDI-Device Identifier10827002227573
UDI-Public(01)10827002227573(17)241228(10)W4550213
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K920703
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberPEG-24-PULL-I-S
Device Lot NumberW4550213
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/30/2022
Initial Date FDA Received12/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ENDOSCOPE - UNKNOWN MAKE AND MODEL
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient SexFemale
-
-