|
Model Number 4FC12 |
Device Problems
Gas/Air Leak (2946); Material Integrity Problem (2978); Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a cryo ablation procedure, a system notice was received after the start of the second ablation application indicating that the safety system detected fluid in the catheter and stopped the injection.The balloon catheter and coaxial umbilical cable were replaced which resolved the issue.It was then reported that when reinserting the balloon catheter into the sheath, a hemostasis valve defect was suspected.Air was continuously drawn into the syringe during aspiration.The sheath was replaced which resolved the issue. the case was completed with cryo. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Product event summary: the 4fc12 sheath with lot number 0011388878 was returned and analyzed.Visual inspection prior to disassembly and functional testing was performed on the shaft, handle, and dilator.Visual inspection of the shaft area identified a shaft kink/twist at approximately 4 inches from the tip.The performance test with sentinel blackbelt leak tester was performed.The performance test failed.Dissection of the returned device revealed a shaft breach in the handle area.In conclusion, the reported hemostatic valve issue was not confirmed through testing.The sheath failed the returned product inspection due to the shaft kink/twist and shaft breach in the handle.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|