Model Number EL5ML
Device Problem Failure to Fire (2610)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # w7026h.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was received with no damage to the external components.In addition, the tyvek(w7013u) was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, it was noted that the clips were not fed into the jaws in the next actuation of the trigger.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, the feed link was noted broken causing the feeding issues and 15 clips were found inside the clip track.The event reported was confirmed and it is related to improper use of the device.Possible causes for the damage found is due to the jaws may have been restricted during firing, or not fully through the trocar during firing.Please reference the instruction for use for more information.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Event Description
It was reported that during a laparoscopic gallbladder procedure, once the clip applier was placed through the trocar, the clip applier would not allow another clip to slide into the jaws and it locked out.Device replaced and procedure was completed successfully with no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key15998703
MDR Text Key308190070
Report Number3005075853-2022-08716
Device Sequence Number1
Product Code FZP
UDI-Device Identifier30705036001847
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberW7026H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2022
Initial Date Manufacturer Received 12/02/2022
Initial Date FDA Received12/16/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-