Catalog Number 326666 |
Device Problem
Missing Information (4053)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd ultra-fine¿ insulin syringe scale markings were partially missing.The following information was provided by the initial reporter, translated from japanese: "this is a report about a printing issue.According to the customer's report, the scale marking was partially missing.".
|
|
Event Description
|
It was reported that the bd ultra-fine¿ insulin syringe scale markings were partially missing.The following information was provided by the initial reporter, translated from japanese: "this is a report about a printing issue.According to the customer's report, the scale marking was partially missing.".
|
|
Manufacturer Narrative
|
H6.Investigation summary: no samples were returned therefore the investigation was performed based on the photo provided.The customer returned an image of the 0.5ml bd syringe.The customer reported that the scale marking was partially missing.The image was examined, and the phot reveals the scale markings fading/missing on the syringe.Due to the batch being unknown, no dhr review can be completed.Embecta was able to confirm the customer¿s indicated failure based on the photo received (scale marking missing).
|
|