Catalog Number 530.705 |
Device Problems
Output below Specifications (3004); Complete Loss of Power (4015); Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported from thailand that during an unspecified surgical procedure, it was observed that the battery reamer device as reported to have been powerless and the tool could not be used because it didn't have enough power to cut the bone.It was reported that device would run for a while, heat up and then turn off.The device would then cool down some, could be used but couldn't cut the bone.It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi:(b)(4).
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device would not run - electrical control unit damage , the handpiece was not working and failed pre-test for check function of the device and check power with the test bench.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component wear.
|
|
Search Alerts/Recalls
|
|
|