Model Number 6248VI-90 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during the implant procedure, the slitter showed abnormality at the knifepoint in comparison to a new slitter.It was noted that during the slitting process of the catheter, the physician felt resistance which led to stopping of the slitting process about twenty centimeters before the end of the catheter.The physician cut the last wat step by step with the same slitter and with a scissor.Afterwards, inspection of the slitter showed abnormality at the knife point.It was also noted that the wrench was broken during control fixation of the right atrial (ra) lead.The wrench was used to connect the leads to the device.It was noted that due to artifacts on the ra lead channel, the physician loosens the ra lead again and screwed it in again.No artifacts were visible anymore.Afterwards the physician used the screwdriver again to check the already screwed in ra lead, only for safety reasons, however the screwdriver was broken without a visible reason.All leads were screwed in, and fixation was verified by the pul l test.No patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|
|
|