Catalog Number 07671687003 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The returned test strips were measured with the returned meter in comparison with the current test strip master lot.For this purpose, two human blood samples from marcumar donors and internal reference meters were used.Test 1: 2.5 inr.Test 2: 2.1inr.All inr values were within the specified target ranges, confirming the functionality of the coaguchek measuring system.No error messages occurred.The returned customer material and retention material comply with the specification.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Occupation was patient/consumer.
|
|
Event Description
|
There was an allegation of questionable results from coaguchek inrange meter serial number (b)(4).At 09:20, the result from the meter was 4.3 inr.At 09:30, the result from a laboratory using recombiplastin reagent was 2.88 inr.At 09:40, the result from an inrange meter at the doctor's laboratory was 3.1 inr.The therapeutic range was 2-3 inr and the patient tests once a week.
|
|
Search Alerts/Recalls
|
|