ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ZEPHYR XL DR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5826
Device Problem No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the pacemaker was failing to pace after the leads were connected.The pacemaker was not used.The physician continued to use the second pacemaker and completed the procedure.There were no patient consequences reported.
 
Manufacturer Narrative
The reported event of ¿no pacing¿ could not be confirmed.As received, the device was above elective replacement indicator level.The results of all electrical test performed, including mechanical stress testing, battery assessment and output verification indicate normal device characteristics.All header connector measurements are within the standards.Longevity assessment was performed, and device was in normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR XL DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16012084
MDR Text Key306003896
Report Number2017865-2022-49084
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model Number5826
Device Catalogue Number5826
Device Lot NumberP000133762
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2022
Initial Date FDA Received12/20/2022
Supplement Dates Manufacturer Received02/28/2023
Supplement Dates FDA Received03/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-