C.R. BARD, INC. (BASD) -3006260740 GLIDEPATH; DIALYSIS CATHETER Back to Search Results
Model Number 5393230
Device Problems Restricted Flow rate (1248); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 12/2022).
 
Event Description
It was reported through results of a clinical trial that approximately two months post hemodialysis catheter placement procedure, the subject experienced an adverse event of catheter occlusion causing decreased catheter flow and high venous pressure.It was further reported that the catheter was removed and exchanged successfully.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 12/2022).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
 
Event Description
It was reported through results of a clinical trial that approximately two months post hemodialysis catheter placement procedure, the subject experienced an adverse event of catheter occlusion causing decreased catheter flow and high venous pressure.It was further reported that the catheter was removed and exchanged successfully.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDEPATH
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16013580
MDR Text Key308117055
Report Number3006260740-2022-05783
Device Sequence Number1
Product Code MSD
UDI-Device Identifier00801741012181
UDI-Public(01)00801741012181
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5393230
Device Catalogue Number5393230
Device Lot NumberREFP2237
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
Patient Weight79 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-