• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS LUCERA DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS LUCERA DUODENOVIDEOSCOPE Back to Search Results
Model Number JF-260V
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported that the brush would not pass.There was no report of patient or user injury due to the event.The subject device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, it was observed that the forceps could not be inserted or removed.This report is being submitted for the reportable malfunction of forceps could not be inserted or removed.
 
Manufacturer Narrative
This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.It has been over 12 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, although a conclusive root cause could not be determined, it is likely the forceps could not be inserted or removed due to following: - the event occurred due to deformation of the forceps channel.- the event occurred due to breakage of the instrument.- the event occurred due to foreign material in the forceps channel.The following information is stated in the ifu (instructions for use) which may help to prevent the issue: operation manual.3.8 inspection of the endoscopic system.- inspection of the instrument channel and forceps elevator.Operation manual.Important information ¿ please read before use.- precautions.Operation manual.3.6 inspection of ancillary equipment.Operation manual.4.3 using endotherapy accessories.Reprocessing manual.5 reprocessing the endoscope (and related reprocessing accessories).Reprocessing manual.7 reprocessing endoscopes and accessories using an aer/wd.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16015283
MDR Text Key308247149
Report Number9610595-2022-05755
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberJF-260V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2022
Was the Report Sent to FDA? No
Date Manufacturer Received01/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-