MIDMARK CORPORATION RITTER EXAM TABLE; DEVICE, MEDICAL EXAMINATION, AC POWERED Back to Search Results
Model Number 104
Device Problem Temperature Problem (3022)
Patient Problem Superficial (First Degree) Burn (2685)
Event Date 12/01/2022
Event Type  malfunction  
Event Description
Patient received a burn when a vaginal speculum was removed from healing drawer in exam table.Patient quickly alerted provider of warmth of instrument and instrument was removed.Table was turned off and equipment removed.The exam table is a ritter model 104 purchased in feb 2001, asset # (b)(4).Tables are inspected annually, next due for inspection feb 2023.The temperature of drawer was reported to be 200 degrees f.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RITTER EXAM TABLE
Type of Device
DEVICE, MEDICAL EXAMINATION, AC POWERED
Manufacturer (Section D)
MIDMARK CORPORATION
MDR Report Key16016074
MDR Text Key305980623
Report NumberMW5113861
Device Sequence Number1
Product Code KZF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number104
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/19/2022
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age35 YR
Patient SexFemale
-
-