• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8603800
Device Problems Gas Output Problem (1266); Failure to Deliver (2338); Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2022
Event Type  malfunction  
Event Description
It was reported that a vent fail occurred during use.There was no patient injury reported.
 
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
It was reported that a vent fail occurred during use.There was no patient injury reported.
 
Manufacturer Narrative
The log file analysis revealed that during the case in question the device detected an internal communication error onboard a pcb that controls the communication between user interface, gas mixer and ventilator.The workstation initiated an emergency shutdown of automatic ventilation and posted corresponding alarms.Monitoring and manual ventilation remain possible in this state.The general issue is known from earlier occurrences of the same phenomenon.Intensive evaluation of the provided information and testing of concerned materials did not result in any finding.The fact that the involved pcb is used in another functional unit of the entire device without showing similar symptoms makes a design problem rather unlikely.A reasonable explanation would be emc disturbances beyond the immunity barriers of the workstation which is compliant to iec 60601-1-2.Therefore, it is recommended that the conditions at the user facility should be checked to prevent from emc disturbance.The number of similar cases, related to the same phenomenon, is within the expected range of the respective risk assessment and thus accepted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
MDR Report Key16016588
MDR Text Key306721461
Report Number9611500-2022-00349
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8603800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/31/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-