• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC IS TABLE TOP ROD CUTTER; INSTRUMENT, CUTTING, ORTHOPEDIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC IS TABLE TOP ROD CUTTER; INSTRUMENT, CUTTING, ORTHOPEDIC Back to Search Results
Model Number 205030
Device Problems Break (1069); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Device report from depuy synthes reports an event in xxx as follows: depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Reporter is a j&j sales representative.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date the table top rod cuter would not cut the rod and broke.A second cutter was retrieved, therefore, no impact to the case and no surgical delay.This complaint involves one(1) device.This report is for one (1) is table top rod cutter.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: device returned.H4: manufacture date added.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that is table top rod cutter, had the cutters are loose.No other problems were found.A dimensional inspection was unable to be performed for the is table top rod cutterr due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the is table top rod cutter would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Dimensional inspection: n/a.Device history lot part # 205030.Lot # 0908mi.Supplier = (b)(4).Batch1: lot qty of (b)(4) unit were released on 09 oct 2008 with no discrepancies.No ncrs were generated during production.Device history review the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.D10: concomitant device added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IS TABLE TOP ROD CUTTER
Type of Device
INSTRUMENT, CUTTING, ORTHOPEDIC
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key16017907
MDR Text Key308155946
Report Number1526439-2022-02181
Device Sequence Number1
Product Code HTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number205030
Device Catalogue Number205030
Device Lot Number0908MI
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNKNOWN RODS
-
-