Catalog Number 306546
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: it was reported the plunger was sticking.As a sample was unavailable for return, a thorough sample investigation could not be completed.It could be possible that the customer is getting some products that are towards the high specification limit for plunger movement and are related to the symptom reported by the customer since they require extra force than normal to expel the solution.A device history record review was completed for provided material number 306546, lot number 2227080.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.Previous investigations have revealed that improper silicone application within the syringe barrel can create plunger resistance.Several quality initiatives have been implemented on our manufacturing line to ensure that the silicone application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the silicone is uniformly applied to the syringe barrel.Based on the investigation and with no sample analysis the symptom reported by the customer could not be confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that 3 bd posiflush¿ normal saline syringes had difficult plunger movement during use.The following information was provided by the initial reporter: "we were having issues with some of the 10ml saline flushes from bd.The plunger has been sticking about halfway through pushing it.It mimics the feeling that your iv has been infiltrated and could potentially cause someone to re-start an iv.There have been at least 3 instances of this occurring.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16018150
MDR Text Key308239786
Report Number1911916-2022-00766
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K213955
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number306546
Device Lot Number2227080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2022
Initial Date FDA Received12/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-