• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; PRL / ILIAC COVERED STENT, ARTERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; PRL / ILIAC COVERED STENT, ARTERIAL Back to Search Results
Catalog Number BXA082902E
Device Problems Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682)
Event Date 06/09/2020
Event Type  Injury  
Manufacturer Narrative
Heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.(b)(4).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
It was reported to gore, that on (b)(6) 2020, the patient presented with a thoracoabdominal aneurysm type ii that was treated with a customized fenestrated endograft (cook medical) and three gore® viabahn® vbx balloon expandable endoprostheses (vbx-device) as branched devices in the celiac trunk and renal arteries.On (b)(6) 2020, a type ic endoleak in the right renal artery was treated with the implantation of an additional stent.The cause of the endoleak was mentioned to be undersizing of the vbx-device.No aneurysm growth was reported.The patient tolerated the procedure.
 
Manufacturer Narrative
Section h6: updated codes for "investigation findings" and "investigation conclusion" due to completed investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Type of Device
PRL / ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
katharina herkt
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key16022061
MDR Text Key305881933
Report Number2017233-2022-03592
Device Sequence Number1
Product Code PRL
Combination Product (y/n)Y
Reporter Country CodeDA
PMA/PMN Number
P160021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/12/2022
Device Catalogue NumberBXA082902E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2022
Initial Date FDA Received12/21/2022
Supplement Dates Manufacturer Received02/24/2023
Supplement Dates FDA Received04/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/13/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age47 YR
Patient SexMale
Patient Weight99 KG
-
-