OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device had no power.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Is unknown, no information provided to date.Udi number is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
One warmer device was received for investigation.During visual inspection the device was observed to have a cracked enclosure and tank cover, faulty speaker, and worn line cord.The device was functionally tested, however the reported issue could not be reproduced, the warmer powered on as intended.During the testing, the investigation did identify an unrelated problem with the device speakers not functioning when alarms were activated.Based on the results of the testing, the investigation could not confirm the reported complaint.No root cause could be determined for the reported power issue, or the observed speaker issue.As no manufacturing root cause could be identified, no review of manufacturing device history records was conducted.A review of device confirmed this device has not been in for service in the previous year and there are no indications of a service issue.Due to the age of the device, it has been deemed beyond economical repair and will be decommissioned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16024371
MDR Text Key307797340
Report Number3012307300-2022-28074
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/21/2022
Supplement Dates Manufacturer Received04/03/2023
Supplement Dates FDA Received04/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/12/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-