Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of surgical revision for infection cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition as associated with the interaction between the implanted device and the patient due to patient illness, unique anatomy, or other condition that impacts the performance of the device.Concomitant device(s): 6700700, 320-01-42, equinoxe reverse 42mm glenosphere, 6746350, 320-15-05, eq rev locking screw, 6727094, 320-10-00, equinoxe reverse tray adapter plate tray +0, 4472341, 304-21-09, 8.5mm platform fx stem left, 6738026, 320-20-00, eq reverse torque defining screw kit.
|
As reported, approximately 1.5 years post op the initial left reverse total shoulder arthroplasty, this 68 y/o male patient was revised due to infection.The patient was revised to exactech devices.The event is not associated with breakage of device & did not lead to surgical delay/prolongation.The reported final outcome for this case is that the patient required prolonged hospitalization as a direct result of this issue.The device will not return, disposed of by facility.No additional information.
|