Reportable based on device analysis completed on 02dec2022.It was reported that the guidezilla encountered crossing difficulties.The 73% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.A 6f guidezilla guide extension catheter was selected for use.During procedure, it was noted that the device could not cross the lesion.The procedure was completed with another of the same device.No complications were reported and the patient was stable post procedure.However, device analysis revealed partial collar and shaft detachment.
|
Device evaluated by mfr: the complaint device was received for product analysis.The device was visually and microscopically examined.There was blood present in the shaft of the device.There was partial collar and shaft detachment.At 17.8cm from the tip of the device, the shaft, markerband and tip were flattened.Angio media provided depicted no damage.Product analysis could not confirm the reported event, as the clinical circumstances could not be replicated.The flattened shaft, tip and markerband, as well as partial collar and shaft detachment found to the device during analysis could contribute to the crossing difficulty which would likely cause resistance upon entry and advancement of the device.No other issues were identified during the product analysis.
|