MAUDE Adverse Event Report: STERIS CANADA ULC RELIANCE VISION SINGLE CHAMBER WASHER/DISINFECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS CANADA ULC RELIANCE VISION SINGLE CHAMBER WASHER/DISINFECTOR Back to Search Results
Device Problem Mechanical Jam (2983)
Patient Problem Insufficient Information (4580)
Event Date 12/02/2022
Event Type  malfunction  
Event Description
The user facility reported an obstruction in the reliance vision single-chamber washer disinfector which caused a rack to jam in the washer door.An employee attempted to manually adjust an instrument on the rack to clear the obstruction.When the obstruction was cleared from the door the washer's door quickly closed and contacted the employee's hand.The employee went to the er and was administered medical treatment and missed a few days of work as a result of the injury.
 
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the washer.The technician inspected the washer, found no issue with the function of the unit or its door obstruction safety mechanism on the load and unload sides.There were no repairs/adjustments required, the unit was tested, confirmed to be operating according to specification, and returned to service.Additionally, the door obstruction detection assemblies were tested during the last preventive maintenance 9/14/2022, and they were found to conform to specifications.The reported event is attributed to user error as the employee should not have stuck their hand in the washer as the door was closing.The reliance vision single chamber washer operator manual (pg.1-1) states, "warning - personal injury and/or equipment damage hazard: if an obstruction is present in the chamber door, obstruction sensor will detect obstruction and door will not close.Wait until water flow has stopped before removing an obstruction." the operator manual further states, "warning personal injury hazard risk of pinch point between door and upper panel.Do not push on top portion of doors; do not push on door when door is rising; do not push on door when door is jammed." the service technician counseled o the user facility personnel on safe operating protocols of the reliance vision single chamber washer.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE VISION SINGLE CHAMBER WASHER/DISINFECTOR
Type of Device
WASHER/DISINFECTOR
Manufacturer (Section D)
STERIS CANADA ULC
490 boulevard armand-paris
quebec, G1C 8 A3
CA  G1C 8A3
Manufacturer (Section G)
STERIS CANADA ULC
490 boulevard armand-paris
quebec, G1C 8 A3
CA   G1C 8A3
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key16033803
MDR Text Key308394530
Report Number9680353-2022-00032
Device Sequence Number1
Product Code MEC
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/02/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-