Model Number 05.000.008
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2022
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional device product codes: hxx, hwe reporter is a j&j employee.Part # 05.000.008.Synthes lot # 007754.Supplier lot #: 007754.Release to warehouse date: 17 feb 2020.Supplier: (b)(4).No ncr's generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Service and repair evaluation: the customer reported that for the device 05.000.008, handle battery powered driver all speeds barely move.The repair technician reported contact pins were damaged, membrane vent and wires were discolored, barrier was burnt and brown residue were found on barriers, circuit board was damaged, motor was rusted, the device ran low in fast forward, run intermittently in reverse conditions.Preventive maintenance is the reason for repair.The cause of the issue is unknown.The following parts were replaced: circuit board, shrink tubing, motor/gearhead, cam screw, membrane switch/flex circuit and all applicable components.The item was repaired per the inspection sheet, passed synthes final inspection on 23-nov-2022 and will be returned to the customer upon completion of the service and repair process.Service record router completed.Finalized service record will be archived.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2022, during a weekly audit, it was found that on the hand piece for battery powered driver, all speeds barely move.There was no patient involvement.No further information is available.This report involves one hand piece for battery powered driver.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAND PIECE FOR BATTERY POWERED DRIVER
Type of Device
SCREWDRIVER, SKULLPLATE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16034664
MDR Text Key308348321
Report Number8030965-2022-11441
Device Sequence Number1
Product Code GXL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number05.000.008
Device Catalogue Number05.000.008
Device Lot Number007754
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/17/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-