|
Model Number 45007 |
Device Problems
Entrapment of Device (1212); Mechanical Problem (1384); Device Displays Incorrect Message (2591); Air/Gas in Device (4062)
|
Patient Problem
Pseudoaneurysm (2605)
|
Event Date 12/02/2022 |
Event Type
Injury
|
Event Description
|
It was reported that a pseudoaneurysm occurred after atherectomy.A 2.1mm jetstream xc catheter was selected for use in an atherectomy procedure to treat a total occlusion.The vessel was mildly tortuous.The catheter made several passes through the lesion.While attempting another pass through the lesion, in blades up mode, it was noted there was difficulty advancing the device.The physician attempted to rex the device backwards.However, a bubl and tube error occurred.They attempted to restart the entire system, but it was unsuccessful.The catheter was attempted to be removed; however, it was stuck on the non-bsc guidewire.The guidewire and catheter were removed together.After atherectomy, a pseudoaneurysm was observed.A non-bsc stent was placed on december 12 to treat the pseudoaneurysm.There were no further patient complications.
|
|
Manufacturer Narrative
|
Device eval by manufacturer: the returned product that was received consisted of a jetstream xc-2.1 atherectomy catheter.No guidewire was returned in or with the device.The device was visually and microscopically inspected for damage.The shaft showed no damage.The set up was conducted using the instructions per the instructions for use.The device primed as designed.The device functioned as designed for a period of 2 minutes with no issues or errors.The setup was repeated multiple times with the same outcome, no bubble, or tubing errors.A.014 thruway guidewire was attempted to be inserted into the devices tip.The wire passed into the device approximately 4cm and stopped.The devices tip was dissected to verify any reason that the wire would not transcend.No obstruction was noticed.The shaft was dissected, and it was noticed that there was clot and coating from the guidewire occluding the guidewire lumen.Inspection of the remainder of the device revealed no damage or irregularities.The complaint for a bubble/tubing errors was not confirmed; however, guidewire sticking was confirmed.
|
|
Event Description
|
It was reported that a pseudoaneurysm occurred after atherectomy.A 2.1mm jetstream xc catheter was selected for use in an atherectomy procedure to treat a total occlusion.The vessel was mildly tortuous.The catheter made several passes through the lesion.While attempting another pass through the lesion, in blades up mode, it was noted there was difficulty advancing the device.The physician attempted to rex the device backwards.However, a bubble and tube error occurred.They attempted to restart the entire system, but it was unsuccessful.The catheter was attempted to be removed; however, it was stuck on the non-bsc guidewire.The guidewire and catheter were removed together.After atherectomy, a pseudoaneurysm was observed.A non-bsc stent was placed on december 12 to treat the pseudoaneurysm.There were no further patient complications.
|
|
Search Alerts/Recalls
|
|
|