ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD Back to Search Results
Model Number MN10450-50A
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2022
Event Type  Injury  
Event Description
It was reported that during a lead revision a lead was stuck in the ipg header, when removing the lead from the epidural space the lead kinked and the contacts were sheared off in the epidural space.Surgical intervention took place where the leads were explanted and replaced.Fragments were unretrievable and remain in the epidural space.Surgical intervention may take place at a future date to address the issue.Investigation was unable to determine which of the lead attributed to the event.
 
Manufacturer Narrative
Date of event estimated.During processing of this incident, attempts were made to obtain complete patient information.Further information was requested but not received.Additional components potentially involved in the event include: common device name: lead, model: mn10450-50a, udi: (b)(4), serial: (b)(4), batch: 7278915.
 
Manufacturer Narrative
It was reported that during a lead revision a lead was stuck in the ipg header, when removing the lead from the epidural space the lead kinked and the contacts were sheared off in the epidural space.Surgical intervention took place where the leads were explanted and replaced.Fragments were unretrievable and remain in the epidural space.Surgical intervention may take place at a future date to address the issue.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Type of Device
DRG SLIM TIP LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key16036203
MDR Text Key306018153
Report Number1627487-2022-07008
Device Sequence Number1
Product Code PMP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/17/2021
Device Model NumberMN10450-50A
Device Catalogue NumberMN10450-50A
Device Lot Number7278915
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2022
Initial Date FDA Received12/22/2022
Supplement Dates Manufacturer Received03/03/2023
Supplement Dates FDA Received03/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DRG LEAD (X2)
Patient Outcome(s) Other;
Patient SexFemale
-
-