Model Number 305916 |
Device Problem
Complete Blockage (1094)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/05/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd safetyglide¿ needle was clogged.The following information was provided by the initial reporter: can you please provide additional details in regards to the issue? the needles wont inject any fluid like the needle is clogged up did you experience any adverse events as a result of this reported defect? no adverse events.
|
|
Manufacturer Narrative
|
H6: investigation summary: no samples or photos received by our quality team for evaluation.A device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Based on the quality team¿s investigation, the root cause of the incident clogged needle can be traced to the manufacturing process.During review of the process, process variations during the needle lubricant application can create clogged needles.Several quality initiatives have been implemented on our manufacturing line to ensure that the needle lubricant application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the needle lubricant is applied uniformly to the needle.
|
|
Event Description
|
It was reported that the bd safetyglide¿ needle was clogged.The following information was provided by the initial reporter: can you please provide additional details in regards to the issue? the needles wont inject any fluid like the needle is clogged up did you experience any adverse events as a result of this reported defect? no adverse events.
|
|