Model Number SONATA |
Device Problems
Impedance Problem (2950); Output Problem (3005)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 12/20/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The user had suddenly no hearing sensations with the device after a trauma occurred.Further technical checks were scheduled.
|
|
Event Description
|
The user had suddenly no hearing sensations with the device after a trauma occurred.The user has been re-implanted.
|
|
Manufacturer Narrative
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress caused by the reported trauma appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
|
|
Manufacturer Narrative
|
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
|
Event Description
|
The user had suddenly no hearing sensations with the device after a trauma occurred.The user has been re-implanted.
|
|