Model Number SONATA |
Device Problems
Impedance Problem (2950); Output Problem (3005)
|
Patient Problem
Failure of Implant (1924)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The hearing performance with the device is affected and in situ testing showed affected channels.Ct scan is considered to check if the electrode is inside the cochlear.
|
|
Manufacturer Narrative
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.No information on possible further steps has been received.
|
|
Event Description
|
The hearing performance with the device was affected and in situ testing showed affected channels.The clinic will discuss how to proceed.
|
|
Manufacturer Narrative
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
|
Event Description
|
The hearing performance with the device was affected and in situ testing showed affected channels.The user has been re-implanted.
|
|