• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problems Mechanical Problem (1384); Failure to Form Staple (2579); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2022
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure, an unexpected big sound was heard, but the clip was closed properly.Then 2 clips came out at same time.One was closed on the target tissue properly, and one was not closed properly.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Batch # x9400p.Additional information was requested and the following was obtained: "did device fire malformed clips? => yes, one was malformed.Two clips were closed properly.Did device fire scissored clips? if other, please clarify =>yes, device fire one scissored clip." attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4) date sent: 1/24/2023 d4: batch # x9400p investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned without apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 8 conforming clips; in addition, the lockout mechanism was found to be non-functional.In order to evaluate the device¿s internal components the instrument was disassembled.Upon disassembling of the device, the ratchet pawl was found damaged causing the lockout failures.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: prior to loading a clip in the jaws and firing the instrument: ensure that the jaws are fully open by verifying that the line of demarcation between the jaws and the instrument shaft is past the distal end of the trocar cannula.Prior to positioning the jaws around the tubular structure or vessel, load a clip into the jaws by partially squeezing the trigger in a smooth continuous motion for approximately one-third of the total firing stroke.Position the jaws with the preloaded clip completely around the tubular structure or vessel to be ligated.The structure to be ligated should be positioned against the apex of the clip.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key16042792
MDR Text Key308459960
Report Number3005075853-2022-08960
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX9400P
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/08/2022
Initial Date FDA Received12/23/2022
Supplement Dates Manufacturer Received01/03/2023
Supplement Dates FDA Received01/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-