Model Number EL5ML |
Device Problems
Mechanical Problem (1384); Failure to Form Staple (2579); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during an unknown procedure, an unexpected big sound was heard, but the clip was closed properly.Then 2 clips came out at same time.One was closed on the target tissue properly, and one was not closed properly.Another device was used to complete the case.There were no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
(b)(4).Batch # x9400p.Additional information was requested and the following was obtained: "did device fire malformed clips? => yes, one was malformed.Two clips were closed properly.Did device fire scissored clips? if other, please clarify =>yes, device fire one scissored clip." attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
(b)(4) date sent: 1/24/2023 d4: batch # x9400p investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned without apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 8 conforming clips; in addition, the lockout mechanism was found to be non-functional.In order to evaluate the device¿s internal components the instrument was disassembled.Upon disassembling of the device, the ratchet pawl was found damaged causing the lockout failures.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: prior to loading a clip in the jaws and firing the instrument: ensure that the jaws are fully open by verifying that the line of demarcation between the jaws and the instrument shaft is past the distal end of the trocar cannula.Prior to positioning the jaws around the tubular structure or vessel, load a clip into the jaws by partially squeezing the trigger in a smooth continuous motion for approximately one-third of the total firing stroke.Position the jaws with the preloaded clip completely around the tubular structure or vessel to be ligated.The structure to be ligated should be positioned against the apex of the clip.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|