VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number PN-004 075 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
During an atrial fibrillation surgery, it was found that the optical fiber of the tacticath quartz ablation catheter was flooded, and the water flowed into the pressure module.After the replacement of the catheter, the surgery was carried out normally and the patient had no adverse reactions.
|
|
Manufacturer Narrative
|
Additional information: d9, g3, g6, h2, h3, h6.One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.Fluid was noted outside of the irrigation tubing but within the tygon tubing proximal to the catheter handle.Further testing revealed a leak at the luer/irrigation tubing junction.The leak at the luer/irrigation tubing junction was determined to be supplier related.
|
|