Model Number 10-85-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a s5 double head pump gave an error message related to motor controller failed and had black screen.There was no patient involvement.
 
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and reported issue could not be reproduced.As a preventive measure, the involved pump was replaced.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
No patient was affected and the problem occurred each time the pump was started.The pump was sent back to manufacturer for further investigation.At the manufacturer site, the error was confirmed and solved by replacing defective computer board (sp pcba hkr 0325 s5 flashed), together with other components listed for compatibility reasons.The unit was then successfully tested and put back in service.Based on the above facts, it cannot be ruled out that the issue was caused by a defective computer board.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key16049645
MDR Text Key306334805
Report Number9611109-2022-00689
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-