LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-85-00 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that a s5 double head pump gave an error message related to motor controller failed and had black screen.There was no patient involvement.
|
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and reported issue could not be reproduced.As a preventive measure, the involved pump was replaced.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
No patient was affected and the problem occurred each time the pump was started.The pump was sent back to manufacturer for further investigation.At the manufacturer site, the error was confirmed and solved by replacing defective computer board (sp pcba hkr 0325 s5 flashed), together with other components listed for compatibility reasons.The unit was then successfully tested and put back in service.Based on the above facts, it cannot be ruled out that the issue was caused by a defective computer board.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|