• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2022
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.The device was returned in two pieces there was a complete shaft break present at 25 cm distal the end of the strain relief.A visual examination identified that the balloon folds were in a wrapped state and had not been subject to positive pressure.There were no issues noted with the balloon material.All blades were securely bonded and no damage to the blades were noted.All blades were intact within their pads and fully bonded to the balloon material.A visual and tactile examination of the hypotube identified a multiple hypotube kinks.The markerbands and tip and of the device were visually and microscopically examined, and no issues were noted with the device that may have potentially contributed to the complaint incident.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 14dec2022.It was reported that the balloon was kinked.The target lesion was located in the severely tortuous and severely calcified artery.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the balloon was kinked.The device was removed from the patient intact and the procedure was completed with a different device.No patient complications were reported.However, device analysis revealed that a shaft break occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16051802
MDR Text Key308007595
Report Number2124215-2022-54920
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeMY
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/17/2024
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0028694786
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2022
Initial Date FDA Received12/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age85 YR
Patient SexFemale
Patient RaceAsian
-
-