ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625374160
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter: occupation is patient/consumer.The test strips were requested for investigation.The reporter's meter and test strips were provided for investigation where they were tested using retention controls.Testing results (qc range = 4.1 - 6.8 inr): qc 1: 5.7 inr.Qc 2: 5.7 inr.Qc 3: 5.7 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
The customer complained of discrepant inr results with coaguchek xs meter serial number (b)(4).The meter result using an unknown test strip lot was 3.4 inr.The customer tested again using an unknown test strip lot and the meter result was 2.6 inr.Customer repeated the meter test again using test strip lot 62682421 and the result was 3.5 inr.Customer's therapeutic range is 2.5-3.5 inr.The customer tests every two weeks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16055108
MDR Text Key308446301
Report Number1823260-2022-04198
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/31/2024
Device Catalogue Number04625374160
Device Lot Number62682421
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2022
Initial Date Manufacturer Received 12/08/2022
Initial Date FDA Received12/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALENDRONATE FOSAMAX; AMOXICILLIN; CALCIUM; KETOROLAC; LANTOPROST; MAGNESIUM; METOPROLOL; PREDNISONE; SIMBRINZA; TRAMADOL; TYLENOL; VITAMIN D; WARFARIN
Patient Age84 YR
Patient SexFemale
-
-