OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR Back to Search Results
Model Number WB91051W
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Olympus was informed by the biomedical technician that the customer high frequency electro-surgical generator unit was getting error messages, ¿e006 insufficient conductivity¿ during a procedure.No death or injury was reported to olympus.
 
Manufacturer Narrative
The concerned unit was returned to the olympus repair center.During the repair inspection/testing, the customer¿s reported e006 error could not be reproduced.However, the units recorded error log shows the multiple errors including e002 (short circuit instrument electrodes coming in contact with each other), e202 (insufficient ne contact) and e006 (no conductive fluid detected).The inspection noted the unit has an old lvps cable.Also, the software is version is an older type.The repair history shows the unit was last serviced via repair in (b)(6) 2018.The investigation is ongoing.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
 
Event Description
The event occurred during a diagnostic procedure.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Section b5 was corrected with information that was inadvertently omitted from the initial report.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of error e006 could not be determined.Error e006 may be caused by increased electrical resistance between two electrodes due to poorly or insufficiently plugged contacts on the conveyor or connecting cable, defective contacts, or growing tissue deposit on the electrode.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
ELECTROSURGICAL GENERATOR
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16057064
MDR Text Key307703287
Report Number9610773-2022-00748
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Device Catalogue NumberWB91051W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-