• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-04
Device Problems Mechanical Problem (1384); Impedance Problem (2950)
Patient Problems Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/10/2022
Event Type  malfunction  
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Event Description
A review of the recipient's test data indicated impedance issues.Programming adjustments were made, however the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Additional information: section d.9.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed tool damage to the silicone overmold on the top and bottom covers, as well as the magnet pocket.In addition, the electrode was severed.These anomalies are believed to have occurred during revision surgery.The photographic imaging inspection revealed a broken electrode within the electrode pocket.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
jennifer rhudy
28515 westinghouse place
valencia, CA 91355
MDR Report Key16060534
MDR Text Key306408397
Report Number3006556115-2022-02198
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2020
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age6 YR
Patient SexMale
-
-