This will be filed to report a leak and intervention.It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with grade of 4.During the procedure the steerable guide catheter (sgc) could not hold fluid column.Air was in the column and the physician aspirated to remove it.The sgc was replaced, and the procedure was continued.One clip was implanted with no reported issue, reducing mr to grade 1+.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
The reported sgc flush port break was confirmed via returned device analysis.The reported leak/splash (loss of fluid column during procedure) could not be replicated in a testing environment.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.All available information was investigated, and the reported leak is due to the broken sgc flush port.However, a cause for the broken sgc flush port could not be determined.The reported unexpected medical intervention is the result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design, or labeling.
|