MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 LCS HP DIST OUTRIGGER; KNEE INSTRUMENT : ALIGNMENT DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 LCS HP DIST OUTRIGGER; KNEE INSTRUMENT : ALIGNMENT DEVICES Back to Search Results
Model Number 9505-01-065
Device Problems Device-Device Incompatibility (2919); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2022
Event Type  malfunction  
Event Description
The lcs hp distal outrigger jammed.Difficulty disengaging it from the distal cutting block.Mechanism jammed.No adverse affects on patient.
 
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The investigation could not verify or draw any conclusions about the root cause of the reported event without the device to examine.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with this report was returned to depuy synthes for evaluation.Visual inspection cannot confirm, the reported allegation.The device came by itself separated from the mating component.Therefore, the jammed condition cannot be confirmed.Handling of the device found nothing that indicates, a functional malfunction,.A functional test could not be performed, due to the mating component was not returned.Additionally, some stains of unknown substances were found.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation (mre), was not possible, because the date code provided is not a finished goods lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS HP DIST OUTRIGGER
Type of Device
KNEE INSTRUMENT : ALIGNMENT DEVICES
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16061197
MDR Text Key307819797
Report Number1818910-2022-26187
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9505-01-065
Device Catalogue Number950501065
Device Lot NumberA1106
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received01/16/2023
01/31/2023
02/22/2023
Supplement Dates FDA Received01/16/2023
02/02/2023
02/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexFemale
-
-