BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4063K
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter facility name: (b)(6) hospital.Additional initial reporter e-mail: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume folfusor overinfused; further described as was infusing too quickly and finished in 12 hours instead of 24 hours.The device was filled with flucloxacillin 8g plus citrate buffer in 230ml sodium chloride 0.9%.This was observed after use of the device.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was connected midline to the upper arm of the patient and stored over the shoulder during the infusion.The flow restrictor was taped to the skin for the entire infusion.The device was manufactured from july 11, 2022 - july 13, 2022.The actual device was received for evaluation.The sample did not contain fluid in the bladder.A visual inspection on the returned sample via the naked eye showed no signs of physical abnormality that could have caused the reported flow problem.A functional flow rate test was performed on the sample and the results were found to be within the product specification range.The reported condition was not verified.The product met specification.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16062467
MDR Text Key307888454
Report Number1416980-2022-07113
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C4063K
Device Lot Number22G016
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received02/02/2023
Supplement Dates FDA Received02/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CITRATE BUFFER; FLUCLOXACILLIN; SODIUM CHLORIDE
Patient SexFemale
-
-