ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Model Number 1468728 |
Device Problems
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
Patient Problems
Low Blood Pressure/ Hypotension (1914); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
One used list#: 146870489 primary plum set was received for evaluation.A cut was observed in the tubing.The set was tested per product specifications.There was a leak from the cut in the tubing.The manufacturing process was reviewed and there are no sharp objects or instruments on the manufacturing floor capable of the observed damage as a preventative measure.The set was tested per product specifications and a leak was observed from the cut tubing.The probable cause of the cut tubing is a sharp object.How, when, or where the cut happened cannot be determined, however, it would have occurred outside of icu medical possession.Lot history review was performed and no nonconformities were identified that may have contributed to the reported complaint.
|
|
Event Description
|
The event involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch which was reported to have holes found in the tubing while levophed was infusing in the icu during patient use.The hole was observed at a location close to patient connection.There was a drop in the patient¿s blood pressure (into the 40¿s); the tubing was changed and the patient stabilized.The patient required titration of the levophed up for some period afterwards, until the blood pressure recovered.
|
|
Search Alerts/Recalls
|
|
|