• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDDRA500Q
Device Problem Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2022
Event Type  malfunction  
Event Description
It was reported that an error message was observed on the device.Technical support was contacted and the device was not implanted to resolve the event.The patient was stable and will continue to be monitored.
 
Event Description
Additional information was received that this event occurred during an initial implant procedure and the device was replaced to resolve the event.
 
Manufacturer Narrative
The reported event of premature battery depletion was confirmed.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.Analysis of device image revealed the device was drawing high current.Electrical testing confirmed high current drain from the hybrid.The cause of the high current was found to be a hybrid anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16064654
MDR Text Key307197338
Report Number2017865-2022-50938
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDDRA500Q
Device Catalogue NumberCDDRA500Q
Device Lot NumberP000155420
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-