DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Device Contamination with Body Fluid (2317)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2022
Event Type  malfunction  
Event Description
It was reported that during use the cs300 intra-aortic balloon pump (iabp) unit had blood back.The fse recommended that the customer discontinue therapy and notify the physician that the unit had been compromised.The customer noted that the iab was a arrow catheter, but called because the blood had gone into the pump through the helium tubing.The customer stated that they would tag the unit for the biomedical team.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp.The fse was able to confirm the reported blood back.To address the issue the fse installed a new purge valve assembly, blood detect tubing, crm, and a safety disk.All functional and safety tests were passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.(b)(6).This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16065395
MDR Text Key308500858
Report Number2249723-2022-03464
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ARROW IAB CATHETER
-
-