DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Device Contamination with Body Fluid (2317)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/08/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during use the cs300 intra-aortic balloon pump (iabp) unit had blood back.The fse recommended that the customer discontinue therapy and notify the physician that the unit had been compromised.The customer noted that the iab was a arrow catheter, but called because the blood had gone into the pump through the helium tubing.The customer stated that they would tag the unit for the biomedical team.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp.The fse was able to confirm the reported blood back.To address the issue the fse installed a new purge valve assembly, blood detect tubing, crm, and a safety disk.All functional and safety tests were passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.(b)(6).This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
|
|