A getinge service territory manager (stm) was dispatched to investigate.The stm evaluated the iabp unit and observed blood in the drain tubing.The stm replaced the safety disk (0997-00-0985-01), condensation removal module (crm) (0997-00-0986-01), purge valve assembly (0104-00-0026), and the blood detect tubing (0008-00-0312).The stm noted that no other signs of blood were found throughout the iabp unit.Subsequently, the stm completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
|