Model Number 72081-01
Device Problems Device Alarm System (1012); No Audible Alarm (1019)
Patient Problems Hypoglycemia (1912); Shaking/Tremors (2515); Cramp(s) /Muscle Spasm(s) (4521)
Event Date 12/17/2022
Event Type  Injury  
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.N/a was selected as it is unknown if the user was using android, ios, or a reader with the fs libre 3 sensor.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An alarm issue was reported with the abbott diabetes care (adc) device.The low glucose alarm did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced "tremors", "cramps", and a seizure and was unable to self-treat.Customer received unspecified treatment by a non-healthcare professional and was seen at a hospital where they were treated with an unspecified infusion for a diagnosis of hypoglycemia.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An alarm issue was reported with the abbott diabetes care (adc) device.The low glucose alarm did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced "tremors", "cramps", and a seizure and was unable to self-treat.Customer received unspecified treatment by a non-healthcare professional and was seen at a hospital where they were treated with an unspecified infusion for a diagnosis of hypoglycemia.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.Visual inspection has been performed and no issues were observed on sensor patch.Data was downloaded successfully, sensor state 7 with event log 11 and sensor state 8 with event log 9 are an indication that the sensor had terminated due to an error recognized by the sensor.This is a part of software design and is not an indication of product non-conformance.De-cased the returned sensor and performed visual inspection of the printed circuit board assembly (pcba), observed that the molex connector and antenna had been damaged due to de-casing and is unable to be re-soldered/repaired due to the solder pads coming away from the pcba, pqe are unable to perform smu testing without the molex connector connected and antenna.Therefore, this issue is unable to test.Section h6 (investigation findings) code c20 (no findings available) and d15 casue not established was selected, as adc was unable to perform further testing on the returned device.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An alarm issue was reported with the abbott diabetes care (adc) device.The low glucose alarm did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced "tremors", "cramps", and a seizure and was unable to self-treat.Customer received unspecified treatment by a non-healthcare professional and was seen at a hospital where they were treated with an unspecified infusion for a diagnosis of hypoglycemia.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Type of Device
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key16067782
MDR Text Key306331138
Report Number2954323-2022-47510
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/31/2023
Device Model Number72081-01
Device Catalogue Number72081-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 12/21/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received01/04/2023
01/31/2024
Supplement Dates FDA Received01/19/2023
02/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-