Model Number 0475100000 |
Device Problems
Electrical Shorting (2926); Temperature Problem (3022)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/22/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that there was a thermal event.
|
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
|
Manufacturer Narrative
|
Alleged failure: while the biomed department was testing the crossfire 2s, a crossfire console ¿popped¿ salesforce case number (b)(4) the failure(s) identified in the investigation is consistent with the complaint record.The root cause is the rf board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that there was a thermal event.
|
|