Model Number 305916 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/25/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd safetyglide¿ needle was clogged.The following information was provided by the initial reporter: safety needle glide not able squeeze.Seems like there is no hole to pull the vaccine.Not able to push the vaccine through the needle.
|
|
Manufacturer Narrative
|
H6: investigation summary: a device history record review was completed by our quality engineer team for provided material number 305916 and lot number 2010821.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.
|
|
Event Description
|
It was reported that the bd safetyglide¿ needle was clogged.The following information was provided by the initial reporter: safety needle glide not able squeeze.Seems like there is no hole to pull the vaccine.Not able to push the vaccine through the needle.
|
|