A getinge field service engineer (fse) was dispatched to evaluate this unit and replaced the solenoid board, cable clamp, hose barb, reservoir assembly, pneumatic luer, single transducer block assembly, fill manifold, purge valve assembly, blood detect tubing, crm assembly, and safety disk that were effected by the blood back.The stm completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
|