• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT Back to Search Results
Model Number FINETUNER ECHO [FINETUNER 2]
Device Problems No Device Output (1435); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
 
Event Description
The finetuner echo was received at service and repair and the repair request form (rrf) stated that there was a mechanical problem.The user reported that it was difficult to remove the battery and the screen was blocked.
 
Event Description
The finetuner echo was received at service and repair and the repair request form (rrf) stated that there was a mechanical problem.The user reported that it was difficult to remove the battery and the screen was blocked.
 
Manufacturer Narrative
Conclusion: according to the information received, a finetuner echo was sent to service repair due to not functioning.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.Also, other components were missing from the circuit board.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key16076657
MDR Text Key306414206
Report Number9710014-2022-00994
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09008737373429
UDI-Public(01)09008737373429
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberFINETUNER ECHO [FINETUNER 2]
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/28/2022
Initial Date FDA Received12/30/2022
Supplement Dates Manufacturer Received12/28/2022
Supplement Dates FDA Received02/13/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexMale
-
-