• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 12°, 4 MM; TELESCOPES, AUTOCLAVABLE (WITHOUT CHANNEL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 12°, 4 MM; TELESCOPES, AUTOCLAVABLE (WITHOUT CHANNEL) Back to Search Results
Model Number A22001A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has already been returned to olympus.However, since the evaluation/investigation is still ongoing, the exact cause of the user's experience and the reported phenomenon has not been determined yet.This report will be updated if additional significant information becomes available or once the evaluation/investigation has been completed.
 
Event Description
Olympus was informed that during preparation for an unspecified procedure the direction pin of the rigid scope was found to be broken.The intended procedure was completed with a similar device and there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for investigation/evaluation but to olympus shanghai (osh), (received at osh on 2022-12-05).The evaluation at osh found the guide pin detached.Furthermore, a lens breakage was determined and fibers of the light guide were found broken.These issued were most likely caused by excessive force and can thus be attributed to user error.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the rigid scope without showing any abnormalities.The case will be closed from olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 12°, 4 MM
Type of Device
TELESCOPES, AUTOCLAVABLE (WITHOUT CHANNEL)
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key16076983
MDR Text Key308532038
Report Number9610773-2022-00817
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22001A
Device Catalogue NumberA22001A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/30/2022
Supplement Dates Manufacturer Received01/30/2023
Supplement Dates FDA Received02/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OLYMPUS WA22367A PASSIVE WORKING ELEMENT
-
-